We assessed the clinical results and complications associated with a new endoprosthetic replacement system (Mutars) used in 250 patients with a malignant bone or soft tissue tumor. The key features of the system are its cementless, hexagonal-shaped stem (titanium alloy), the possibility of torsion adjustments in 5 degrees -increments, and the Trevira tube for soft tissue attachment. The mean age of the patients was 30.7 years, and the mean followup was 45 months. Prosthetic survival at 5 years was 89.7% for the upper extremity and 68.5% for the lower extremity. Prosthetic survival without any reoperation was 73.4% at 3 years postoperatively and 60.4% at 5 years postoperatively. Prosthetic failure was caused by deep infection in 12% (30 patients) of patients and aseptic loosening in 8% (20 patients) of patients. Stem fracture occurred in only 1.6% (four patients) of patients. Dislocation rates were reduced by using the Trevira tube. Limb survival was achieved in 82.6% to 93.1% of patients depending on the endoprosthetic replacement site, and functional results ranged between 63% to 83% according to the Tumor Society score. Our results suggest limb salvage with the Mutars endoprosthesis is successful with good functional results.Therapeutic study, Level IV (case series).
Support the authors with ResearchCoin