Background: Free flap monitoring is more difficult in patients with dark skin because ischemia and congestion can be masked by pigmentation. For this reason, adjunct methods such as cutaneous near-infrared spectroscopy are of elevated importance in patients with highly pigmented skin. The purpose of this experiment is to determine if ViOpitx T.Ox performance is affected by cutaneous pigmentation. Methods: Swine with naturally occurring areas of nonpigmented and pigmented skin were used. Pigmentation of each animal was assessed using spectrophotometry and histopathology. During normoxemia, tissue oxygenation (StO 2 ) measurements were taken of nonpigmented and pigmented skin using the T.Ox device. A bicolor pedicled rectus abdominis myocutaneous flap was raised, and T.Ox probe was adhered to adjacent areas of opposite coloration on the same flap. StO 2 was measured continuously during reversible episodes of flap ischemia and congestion (n = 4 swine, n = 6 flaps). Results: There was not a significant difference between baseline StO 2 values of nonpigmented (49% ± 7.9%) and pigmented skin (47% ± 6.2%). The absolute change in StO 2 was significantly larger during both ischemia (6%) and congestion (16%) in nonpigmented skin compared with adjacent pigmented skin. Conclusions: T.Ox detects flap ischemia and congestion in both highly pigmented and nonpigmented skin. However, surgeons need to be aware that StO 2 changes related to complete flap ischemia or congestion may be much more subtle than what is seen in nonpigmented skin. This study establishes a novel internally controlled porcine model that isolates the impact of skin pigmentation when assessing cutaneous devices measuring tissue oxygenation.
This paper's license is marked as closed access or non-commercial and cannot be viewed on ResearchHub. Visit the paper's external site.