Utility: This is a quick assay performed to quantify the concentration of total protein in samples. This assay is traditionally used to measure protein concentrations in cell lysates so they can be normalized to each other for western blotting. This is useful because in western blotting you want to load the same protein content from each sample so it's simpler to measure relative expression of your protein of interest.
How it Works: Peptide bonds in proteins of your samples reduce Cu2+ ions from the copper(II) sulfate reagent to Cu1+, then two molecules of bicinchoninic acid form a complex with each Cu1+ ion which yields a purple complex that absorbs light at 562 nm. The assay is measured by passing this light through the samples and measuring how much was absorbed.
Watch Out For: This assay functions via a reduction mechanism, so inclusion of any reagents in your sample such as DTT or BME would interfere with the reaction. It's best to add these after performing this assay, when preparing your samples for SDS-PAGE. If these reagents are present they can be removed via Zeba desalting columns.
Summary: This protocol is re-worded from that of the manufacturer (Thermo Fisher) and includes an updated standard curve preparation scheme. This was done since the initial protocol yielded different final volumes for each standard, which was problematic because this assay uses the same volume of every standard. With this updated dilution series you won't end up with 1 standard that is used up while having 7 others which still have residual volume.
Quick Pro Tips:
Photo: This is what the assay looks like when it's finished. Standard curve is on the right, and the gradient from dark purple to clear shows you a dilution series was made.
Diluent should be the same liquid that your samples are comprised of. This is usually your lysis buffer + protease or phosphatase inhibitors which you used. These standards should be prepared in 1.5 mL tubes and stored -20C before usage. This series is starting from a 2,000 ug/mL stock, which comes standard in this kit.
[Protein] ug/mL | Diluent (uL) | Volume Added | Of | Total Volume |
2000 | 0 | 0 | N/A | 229 |
1500 | 56.50 | 169.50 | 2000 ug/mL standard | 226 |
1000 | 298.0 | 298.0 | 2000 ug/mL standard | 596 |
750 | 56.88 | 170.63 | 1000 ug/mL standard | 227.5 |
500 | 113.75 | 113.75 | 1000 ug/mL standard | 227.5 |
250 | 170.63 | 56.88 | 1000 ug/mL standard | 227.5 |
125 | 199.06 | 28.44 | 1000 ug/mL standard | 227.5 |
0 | 227.5 |
|
| 227.5 |
Most plate readers will have an accompanying software which will construct a standard curve for you. I've included one below for reference:
And these are the standard curve fitting details.
Curve Name | Curve Formula | A | B | R^2 |
StdCurve | Y=A*X+B | 0.26 | 0.0493 | 0.997 |
This one had a particularly good R^2 of 0.997. R^2 is a measure of how well correlated your standard curve is. An R^2 of greater than 0.95 is acceptable. If not, it's recommended to repeat the assay, and if this happens again, re-prepare a new set of standards.
The plate reader will use the standard curve to calculate the concentrations of your samples. It's important here to include any dilution factors your samples may have, and scale them up by that amount. Also, make sure you're using concentration values, not absorbance values.
Once you have your concentrations from the BCA assay, proceed to preparing your samples for running on a gel. Use this calculator I've made to make your sample process much easier: https://www.researchhub.com/post/2034/gel-loading-calculator-for-western-blotting